Achieving good peak shape in HPLC is essential for accurate quantitation, improved resolution, and extended column lifetime. A well-shaped peak is symmetrical, with a tailing factor of ~1.0, high efficiency, and narrow peak width.
Key Factors Influencing Peak Shape
Column Selection & Chemistry
Silica Purity & Acidity:
- High-purity, fully hydroxylated silica reduces peak tailing for basic compounds.
- Bonded Phase & End capping: Double end capped columns (e.g., Eclipse XDB) minimize silanol interactions, improving peak symmetry.
Specialized Phases:
- Bonus-RP (polar-embedded phase) reduces peak tailing for basic analytes.
- Extend-C18 is stable at high pH, preventing ion exchange issues.
- SB-Aq ensures good retention in high aqueous mobile phases.
Mobile Phase Optimization
- pH Control: Using buffered mobile phases prevents peak tailing by stabilizing analyte ionization.
- Organic Modifiers: Acetonitrile vs. methanol can influence peak shape by altering secondary interactions.
- Additives (TEA, TFA, Acetic Acid): Suppress unwanted ion-exchange interactions for basic and acidic compounds.
Injection & Sample Considerations
- Sample Overload: High concentration can cause peak fronting or tailing.
- Injection Solvent Strength: Using a stronger solvent than the mobile phase leads to peak broadening or splitting.
- Metal Contamination: Chelating compounds may interact with metal ions in the system, distorting peaks. Acid washing can mitigate this issue.
Pore Size for Large Molecules
- 300Å wide-pore columns are ideal for peptides & proteins, ensuring sharper peaks by allowing better analyte diffusion.
Best Practices for Optimal Peak Shape
- Choose high-purity, fully hydroxylated silica columns (e.g., Stable Bond, Eclipse XDB, Bonus-RP, Extend-C18).
- Use buffered mobile phases (25–50 mM) to minimize secondary interactions.
- Adjust pH and mobile phase composition to match analyte properties.
- Optimize injection volume, solvent strength, and sample concentration to prevent overload issues.
- Perform acid washes periodically to remove metal contamination.
By following these guidelines, HPLC users can achieve sharper, more reproducible peaks, leading to higher analytical accuracy and longer column lifespan.